To determine the feasibility of transoral endoscopic nasopharyngectomy without division of the soft plate using a flexible, next-generation robotic surgical system. Preclinical anatomic study using four human cadavers. Transoral resection of the nasopharyngeal wall with en-bloc resection of the cartilaginous Eustachian tube and dissection of the parapharyngeal fat space. The first flexible robotic surgical system has recently been described. We performed a series of laboratory experiments to determine whether this flexible system could be used to perform transoral robotic nasopharyngectomy. This novel system allowed docking of the patient-side cart at the side of the operating table. The cannula tip was placed approximately 12 cm from the edge of the retractor pointing superiorly toward the nasopharynx (NP). Retraction of the soft palate anteriorly and tonsillar pillars laterally with stay sutures expanded the velopharyngeal inlet, providing adequate space to deploy all four instruments (three surgical instruments and a camera) into the NP for dissection. All instruments could be deployed into the NP, without collision or restriction of joint movement in this cadaver model. Using this position and docking location, the new flexible surgical robot provided sufficient access, reach, and visualization to complete robotic nasopharyngectomy with en-bloc resection of the cartilaginous Eustachian tube. This feasibility study showed that transoral endoscopic nasopharyngectomy could be performed without compromising the integrity of the soft palate using a novel flexible robotic surgical system. N/A. Laryngoscope, 126:2257-2262, 2016.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call