Abstract

A radiotelemetry transmitter conforming to British Governmental specifications for biomedical telemetry has been devised for either intermittent or continuous recording of intra-arterial pressure in fully ambulant subjects. A1 kHz subcarrier is frequency modulated by the pressure waveform, the subcarrier being in turn frequency modulated on to a crystal-controlled oscillator at 17·060 MHz which, after frequency multiplication, produces an output to drive an integral halfwave aerial at 102·360 MHz. The apparatus has been successfully used for monitoring for periods of 3 to 36 h.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.