Abstract

A transmission-type ionization chamber for monitoring dose power of the Roentgen-TA 150/10 X-ray therapeutic apparatus is described. The X-ray working dose range of the apparatus is 4.5-70 keV. The X-ray chamber sensitivity at Al equivalent 100 nm for X-ray energy 4.5 keV is 2·10–11 A/mGy/sec. Relative spectral sensitivity error of the chamber within the given range is ±13.0 %. The correction of the apparatus reduces this error to ±3.0 %.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.