Abstract

To assess the magnetic field interactions on 60 heart valve prostheses subjected to a 4.7 T MR system. It addresses the question of whether heart valves deemed safe at 1.5 T may pose safety hazards as patients are exposed to increased static magnetic fields. Ex vivo testing was performed to evaluate translational and rotational forces on 60 heart valves using previously described techniques. Translational forces were detected on 58 heart valves ranging from 0.5 degrees to 7.5 degrees. Seven valves exhibited paramagnetic/weakly ferromagnetic behavior, and 51 valves exhibited diamagnetic behavior. Rotational forces were observed for 46 valves. Criteria previously used for safety assessment of heart valve prostheses and expressed in terms of magnetic forces suggest the forces observed in this study are compatible with the safe use of these valves in magnetic resonance (MR) systems with static fields up to 4.7 T.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.