Abstract

To perform long-term comparison between laboratory Stago and Point-of-Care (POC) i-STAT methods for determining the international normalized ratio (INR). This was a multicenter method comparison of patient INR results and factors related to performance variance. For 5 years, the assays demonstrated close patient correlation within and above the 3.5 INR therapeutic range cutoff (bias, 0.23 INR units). Patient results above 3.5 INR were bimodal, with 60% demonstrating an i-STAT INR bias of less than 0.5. Several patient conditions were associated with the presence of a higher i-STAT bias. In year 6, a broader range i-STAT bias developed, increasing to 0.73 INR units. The increased bias persisted for 3 years, then returned to initial levels following i-STAT adjustments. The substantial increase in i-STAT bias after a long period of stability was partly corrected by renewed correlation to the international reference preparation. Additional assay drift is discussed in relation to thromboplastin reagents and other testing variables. This study emphasizes the need for continual laboratory correlation with POC devices and caution in using published comparisons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.