Abstract

At present the United States Department of Agriculture (USDA) Reference Spectroradiometric Network consists of 4 sites: Table Mt. CO, Ft. Collins CO, Lamont OK (The ARM program SGP site), and Beltsville MD. At each site we operate a 1-meter cascaded additive-double Czerny-Turner scanning monochromator with a bi-alkali Photomultiplier and photon-counting detection. Irradiance calibrations are provided for the instrument at Table Mt CO by NOAA's Central Ultraviolet Calibration Facility (CUCF) from NIST-traceable standards. Calibrations are transferred from this instrument to others in the network (and additional stability monitoring of the primary instrument conducted) using shippable transfer calibrators we have designed. Here we describe these transfer calibrators, and our operational experience seen at the Table Mt. Site in 2002 and 2003 to date.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.