Abstract

The ability to transfer calibrations between near-infrared instruments is important for the acceptance and spreading of the technology. A method called transfer by orthogonal projection (TOP) is proposed for deriving calibrations that are robust to between-instrument variation, and thus may be transferred without adjustment. The idea is to orthogonalise the calibration spectra to directions in the spectral space in which most between-instrument variability lies. The method is related to the idea of a repeatability file, but has advantages over that approach when factor-based methods are used for calibration. Two examples are presented. In both of these, a calibration is successfully transferred to instruments not involved at all in its derivation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.