Abstract

Long-term use of intravenous catheters can lead to catheter fracture and embolization of fragments. Transcutaneous retrieval of these catheter fragments can be challenging because of their fragility. We report an 8-year-old boy with Hemophilia disease who underwent removal of intravenous Port catheter after 7 years of use, resulting in embolization of fractured catheter fragments into the distal pulmonary arteries. The snare technique to pull the snared fragment into a sheath was unsuccessful, and it leads to further breakdown due to its fragility. An alternative technique using a combination of a snare kit and a Spider FX™ Embolic Protection Device was employed. This technique allowed the fragments to be secured proximally with the basket device and distally with the snare. The unit was then pulled through a sheath and removed from the body. To our knowledge, Spider FX™ Device has not been used in this way before.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.