Transcranial Doppler (TCD) is an invaluable tool allowing real-time monitoring of physiologic blood flow velocity changes. We present a case where TCD monitoring for vasospasm after subarachnoid hemorrhage identified blood flow velocity changes consistent with sudden increased intracranial pressure (ICP) due to a malfunctioning extraventricular drain. The primary team was alerted to these findings, and immediately revised her shunt with normalization of ICP and TCD. Serial TCD monitoring allowed identification of an imminently fatal complication in time to allow a life saving intervention. TCD is a portable, inexpensive, real-time tool providing important physiologic data regarding blood flow velocities and intracranial pressure that is crucial to the care of critically ill patients.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.