Abstract

We present real-time cavitation monitoring and mapping in a clinical trial for blood-brain barrier (BBB) opening in Alzheimer’s disease (AD) patients using a neuronavigation-guided focused ultrasound (FUS) system. Six AD patients (N = 6, age = 68.5 ± 9.5) were sonicated at the right prefrontal lobe with a single-element FUS transducer (PNP = 0.2 MPa, fc = 0.25 MHz, pulse length = 10 ms, PRF = 2 Hz, duration = 2 min) with the microbubble administration (Definity). We performed passive cavitation detection (PCD) using a single-element hydrophone (N = 4), and passive acoustic mapping (PAM) with a 64-element phased array transducer (N = 2). Five patients showed successful BBB opening (volume = 654 ± 394 mm3) on day0, and closure on day3, confirmed by contrast-enhanced T1-weighted MRI, while one patient served as a negative BBB opening case due to the insufficient microbubble administration. Ultra-harmonic cavitation dose (CD) correlated with the opening volume (R2 = 0.66, N = 4), but harmonic and broadband CDs showed a relatively poor correlation (R2 = 0.06 and 0.33). Cavitation maps (N = 2) showed a spatial distribution of acoustic energy that roughly coincided with the respective BBB opening location. We thus demonstrated the successful BBB opening and cavitation monitoring with the neuronavigation-guided system that allowed for a cost-effective procedure compared to the MR-guided approaches. The PCD and PAM showed promising results for potentially predicting the BBB opening volume and location found in MRI.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call