Abstract

This study assessed the feasibility and efficacy of implanting a new nitinol device in the left atrial appendage (LAA) in a canine model. Left atrial thrombus occurs in the LAA in 90% of humans with nonvalvular atrial fibrillation (AF). The Amplatzer Left Atrial Appendage Occluder Device now known as the Amplatzer Cardiac Plug (Amplatzer Medical LLC, Plymouth MN) is a new transcatheter Nitinol device containing polyester fabric designed to seal the LAA. The device was implanted in 10 canines with immediate, approximately 30 and approximately 90-day follow-up by echocardiography, angiography, and final pathological examination. The device was successfully implanted in all animals and found to be retrievable and repositionable. There was complete occlusion of the LAA in all cases without embolization. There was no thrombus formation on the device and no pericardial effusion or impingement on the circumflex coronary artery, mitral valve, or left pulmonary vein. A single instance of tiny left atrial mural thrombus at device end screw contact was attributed to the small canine left atrium. The success of this animal study confirms safety and feasibility of the Amplatzer Cardiac Plug. Human trials are warranted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.