Abstract

Transcatheter closure of perimembranous ventricular septal defect (PmVSD) is an established procedure. However, the occurrence of complete heart block limits its scope. The newer KONAR-MF™ occluder has specific design characteristics that may improve the safety of PmVSD closure. The objective of the studywas to describe the efficacy and mid-term follow-up of transcatheter closure of PmVSD using KONAR-MF™. The study was conducted prospectively in 3 Indian centers (January 2018-December 2022). PmVSD closure was done by both antegrade and retrograde methods, and patients were followed up at 1, 3, 6, 12months, and annually after that. 121 out of 123 patients were included with the following characteristics: median age 4.4 (0.18-40) years; weight 15 (2.1-88) kg; mean Qp/Qs ratio 1.87 ± 0.52 and pulmonary artery mean pressure: 22 ± 6.9mmHg. The procedure was successful in all but 3; the device was removed due to significant residual shunt (n = 2) and new development of aortic regurgitation (AR) (≥ mild) in 1. The median defect size was 5.2 (2.5-12) mm. Device sizes from 6/4 to 14/12 were deployed (median fluoroscopy time 13.3min; range 3.6-47.8). Shunt occlusion rates were 90%-Immediate, 95%-pre-discharge, and 97%-1month, with no instances of complete heart block after the procedure and during follow-up. Six had new onset AR (mild: 2, trivial 4), and one had increased tricuspid regurgitation. All patients were well during follow-up (median: 20months; range: 6-46). The new KONAR-MF™ occluder appears to be a promising and safe alternative for the closure of the PmVSD; further long-term follow is merited.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call