Abstract

Little is known about the efficacy and safety of the 40 mm Amplatzer septal occluder (ASO). Thirty-three patients (22 female, 11 male) with a large secundum atrial septal defect (ASD) underwent attempted device closure using the 40 mm ASO at a median age of 40 years (range, 14-81 years) and median weight of 65 kg (range, 48-98 kg). The median size of the ASD measured on 2D transesophageal echocardiography (27 patients) or intracardiac echocardiography (6 patients) was 30.5 mm (range, 24-39 mm) and the median balloon-stretched diameter was 37.7 mm (range, 32-43.7 mm). The median Qp:Qs ratio was 3.2:1 (range, 1.4-6.2). The attempt was unsuccessful in five patients; two had device embolization and one had left atrial wall perforation due to the sheath; all three required emergent surgery. The attempt was successful in the 28 remaining patients, resulting in complete immediate closure in 14 and a trivial residual shunt in 14. Fluoroscopy time ranged from 8.6 to 37.8 min (median, 12.2 min). At 24-hr follow-up, 2D transthoracic echocardiography with color flow Doppler revealed complete closure in 23 patients, and 5 had a trivial residual shunt. There were no complications encountered in patients who received the device. On follow-up, all patients are doing well. We conclude that the 40 mm ASO is safe and effective in most patients with a large ASD up to a diameter of 39 mm. However, the use of this device requires careful attention as the procedure may be unsuccessful or the device may embolize.