Abstract
Transatrial placement of a Greenfield filter at the time of cardiac surgery was performed on three patients with recent femoral venous thrombosis and unstable angina, and at the time of pulmonary embolectomy on four others. The filter carrier was passed through the right atrial cannulation site under fluoroscopic guidance into the inferior vena cava. A preliminary venacavogram was performed through the carrier or a previously placed angiographic catheter. Two complications were encountered: the first was misplacement of a filter in the right renal vein before routine venacavography was employed; the second was caudal displacement of a filter by a right atrial venous drainage cannula that projected into the inferior vena cava. The latter complication can be prevented by the use of a right atrial "sump" drain that does not protrude into the inferior vena cava.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.