Abstract
The auto-resonant trap mass spectrometer, ART-MS, utilizes electrostatic ion trapping within an anharmonic potential well. Ions are detected after mass selective trap ejection with auto-resonant driving employing only low-power rf electronics. We identify the major limiting factor in the mass resolution of these instruments. Whilst keeping in the spirit of maintaining a rapid scan rate, low cost, lightweight instrument, with minimal required machining tolerances, we introduce a method for much improving the mass resolutions of an ART-MS. The addition of two electrodes has enabled an improvement in the mass resolution by a factor of ∼4. The scheme significantly reduces the effects of a finite sized trap and compensates for the influence of radial variation in natural oscillating frequencies within the trap. Compensation can be implemented with a wide range of designs and is not limited by the size of the trap.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.