Abstract

To determine and track the planar orientation of active interventional devices without using localizing RF microcoils. An image-based tracking method that determines a device's orientation using projection images was developed. An automated and a manual detection scheme were implemented. The method was demonstrated in an in vivo mesocaval puncture procedure in swine, which required accurate orientation of an active transvascular needle catheter. The plane of the catheter was determined using two projection images. The scan plane was adjusted automatically to follow the catheter plane, and its orientation with respect to a previously acquired target plane was displayed. The algorithm facilitated navigation for a fast and accurate puncture. Using image-based techniques, with no mechanical design changes, the orientation of an active intravascular probe could be tracked.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.