Abstract

This paper reports on the upgrading of a standard solid state NMR spectrometer, which has been used in combination with a field variable 7 T cryomagnet, to a low-cost combined SFG and PFG NMR spectrometer. Both methods are applied to solid lithium as a simple test case. The results show that under the given conditions SFG NMR and PFG NMR can provide tracer diffusion coefficients for 7 Li diffusion down to about 10 −14 and 10 −13 m 2/ s , respectively. SFG and PFG NMR are complementary methods. The paper demonstrates advantages and disadvantages of each method with a concrete example and why it is desirable to be able to apply both methods to the same sample.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.