Abstract
Measurement traceability is the critical aspect in metrology area. As national metrology institute in Indonesia, National Measurement Standard Laboratory-National Standardization Agency of Indonesia, SNSU-BSN, maintains the recommended stabilized laser as primary standard of length, i.e., 633 nm iodine stabilized He-Ne laser (KIM-1) with expanded uncertainty (Ur) 63.4×10−12, was obtained from International Laboratory Comparison as a part of CCL-K11 key comparison of the optical frequency standards. Two calibration methods have been applied to create the traceability link from the length measurement device to KIM-1. The accuracy of KIM-1 transferred to stabilized and non-stabilized He-Ne lasers concerning beat frequency method and direct measurement method using wavelength meter. Laser head Agilent 5519B with dual frequency mode was used as DUT in beat frequency experiment. As calibration result, the optical wavelength of DUT is (632.991 371 061±1.9×10−9) nm and (632.991 374 729±2.6×10−9) nm, respectively for 1st and 2nd polarized beam. In case of direct measurement method, calibration result was obtained by using Laser Head Polytec OPV-503 as DUT, the result is (632.9913±0.0026) nm. Best measurement capability was reached 1.3×10−10 in relative value (Hz/Hz) for beat frequency method and 2.6×10−3 nm for direct measurement method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.