Abstract
In detector output-signal measurements, where the measurement uncertainty critically impacts the total uncertainty of a detector-based radiometric scale realization, traceability of the photocurrent measurement to SI traceable electrical standards is needed. A reference photocurrent-to-voltage conversion scale has been developed and described here. The new standard has internal reference resistors for all signal-gain selections to decrease noise pickup of traditionally used external reference resistors. The internal resistors were calibrated by the NIST Quantum Electrical Metrology Division against standard resistors. Using the substitution method, the reference photocurrent-to-voltage conversion scale was transferred from the converter standard to other converters up to a maximum signal-gain of 1010 V/A with an expanded uncertainty of 0.013% (k=2). The DC conversion scale was extended to AC mode signal-gain calibrations where the photocurrent measurement uncertainty is 0.05% (k-2).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.