Abstract

A method for the determination of trace levels of silicon from biological materials by inductively coupled plasma mass spectrometry (ICP-MS) has been developed. The volatility of water-soluble silicon species, hexafluorosilicic acid (H2SiF6), and sodium metasilicate (Na2SiO3) was investigated by evaporating respective solutions (50 µg/mL silicon) in HNO3), nitric acid + hydrochloric acid (HNO3 + HCl), and nitric acid + hydrochloric acid + hydrofluoric acid (HNO3 + HCl + HF) at 120 °C on a hot-block to near dryness. The loss of silicon from H2SiF6 solutions was substantial (>99%) regardless of the digestion medium. Losses were also substantial (>98%) for metasilicate solutions heated in HNO3 + HCl + HF, while no significant loss occurred in HNO3 or HNO3 + HCl. These results show that H2SiF6 species were highly volatile and potential losses could confound accuracy at trace level determinations by ICP-MS if digestates prepared in HF are heated to eliminate HF. Among the various matrices comprising major elements, sodium appeared to be effective in reducing silicon loss from H2SiF6 solutions. Excess sodium chloride (NaCl) matrix provided better stability, improving silicon recoveries by up to about 80% in evaporated HF digestates of soil and mine waste samples, but losses could not be fully prevented. To safely remove excess acids and circumvent the adverse effects of excess HF (e.g., risk of high Si background signals), a two-step digestion scheme was adopted for the preparation of biological samples containing trace silicon levels. A closed-vessel digestion was performed either in 4 mL of concentrated HNO3 and 1 mL of concentrated HCl or 4 mL of concentrated HNO3, 1 mL of concentrated HCl and 1 mL of concentrated HClO4 on a hot plate at 140 °C. Digestates were then evaporated to incipient dryness at 120 °C to remove the acids. A second closed-vessel digestion was carried out to dissolve silicates in 0.5 mL of concentrated HNO3 and 0.5 mL of concentrated HF at 130 °C. After digestion, digestates were diluted to 10 mL. The solution containing about 5% HNO3 and 5% HF was directly analyzed by ICP-MS equipped with an HF-inert sample introduction system. The limit of detection was about 110 µg/L for 28Si when using the Kinetic Energy Discrimination (KED) mode. The method was used to determine silicon in various plant and tissue certified reference materials. Data were acquired for 28Si using KED and standard (STD) modes, and 74Ge and 103Rh as internal standard elements. There was not any significant difference between the accuracy and precision of the results obtained with 74Ge and 103Rh within the same measurement mode. Precision, calculated as relative standard deviation for four replicate analyses, varied from 5.3 (tomato leaves) to 21% (peach leaves) for plant and from 2.2 (oyster tissue) to 33% (bovine liver) for tissue SRM/CRMs. Poor precision was attributed to material heterogeneity and the large particle size distribution. An analysis of lung tissue samples from those with occupational exposure to silica dust revealed that tissues possessed substantial levels of water-soluble silicates, but the most silicon was present in the particulate matter fraction.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call