Abstract
Ethylene oxide (EtO) is a key carcinogen that is widely used in chemical manufacturing and biotechnology industries. Recent work has suggested that permissible exposure limits for EtO be reduced from 1–5 ppm to sub-ppb levels. Such new standards will require new methodologies that are capable of measuring EtO with the requisite precision. In this paper, we demonstrate a new analyzer based on cavity-enhanced absorption spectrometry that utilizes a broad EtO absorption feature near 3066 cm-1. A fit function is developed that includes water, methane, and EtO absorbances and accounts for absorption both inside and outside the cavity. A methane standard is used to determine the cavity gain factor, and the EtO absorbance spectrum is empirically determined. The final system shows excellent linearity from 0 – 909 ppb EtO (R2 ~0.9999) with a measurement precision of better than ±1 ppb (1s, 60 seconds) that improved to ±0.5 ppb (1s, 15 minutes). Deliberate ambient EtO releases demonstrate the instrument’s utility in rapidly detecting hazardous conditions. Further work will include improving the measurement precision and directly comparing the system to EPA Method TO-15.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.