Abstract

A newly constructed chamber for measuring with MeV proton beams at small incidence angles (0 to 35 mrad) is used to analyse trace elements on flat surfaces such as Si wafers, quartz substrates and perspex. This set-up is constructed in such a way that the X-ray detector can reach very large solid angles, larger than 1 sr. Using these large solid angles in combination with the reduction of bremsstrahlungs background, lower limits of detection (LOD) using TPIXE can be obtained as compared with PIXE in the conventional geometry. Standard solutions are used to determine the LODs obtainable with TPIXE in the optimized setup. These solutions contain traces of As and Sr with concentrations down to 20 ppb in an insulin solution. The limits of detection found are compared with earlier ones obtained with TPIXE in a non optimized setup and with TXRF results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.