Abstract

In this paper, we demonstrated a trace double-component gas detection system based on frequency division multiplexing using two-way distributed feedback laser diode (DFB-LD) as the excitation laser source. In the frequency response range of photoacoustic spectroscopy (PAS) cell, this detection system could detect the second harmonics of different frequencies f0 and f1, which were excited by the gas absorption using the two wavelengths of DFB-LD at modulation frequencies f0/2 and f1/2. Unlike previous double-component gas sensor, these two DFB-LDs connected into the both ends of PAS cell without optical fiber coupler could effectively improve the excited light intensity and simplify the system structure. This feature allowed the trace detection of double-component gas of mixed-methane (CH4) and water (H2O) vapor with a low detection limits 11.055 ppm and 885 ppb, respectively. Additionally, the linear responses to the gas concentration were achieved 0.9986 and 0.9997, and it exhibited potentially as an effective detection system for complex gas component.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.