Abstract

Highly sensitive detection of neutral Kr atoms was accomplished by the use of laser-induced fluorescence. In one experiment, Kr at 40 parts per 10(12) in He was detected at a signal-to-noise ratio of 500 by time-resolved fluorescence measurements. The Kr metastable 1s(5) level was populated by cascade after two-photon excitation to the 2p(6) level by the frequency-tripled output of a pulsed single-longitudinalmode dye laser. After a delay, when scattered laser light and cascade resonance fluorescence became negligible, trace quantities of Kr were detected by the use of a pulsed-laser pumping scheme. In a related experiment, (78)Kr/(86)Kr isotope ratios ranging from 1 to 0.1 were measured with a resonant isotopic depletion technique first proposed by Makarov [Sov. J. Quantum Electron. 13, 722 (1983)]. The (86)Kr metastable population was selectively depleted by optical pumping to a higher-lying state that relaxed to the ground state by means of radiative cascade. After the (78)Kr/(86)Kr ratio of metastables had been enriched by a factor of 10, the (78)Kr population was probed by pulsed excitation. Premixed (78)Kr/(86)Kr ratios were measured to within an accuracy of 10%, even for unresolved, Doppler-broadened transitions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.