Abstract

A cathodic sputtering atom reservoir designed for atomic absorption spectrometry of conducting solid samples is examined as a potential sampling device for trace and ultra-trace detection using laser-induced fluorescence (LIF) spectrometry. The analytical results are promising, with sub-µg g–1 sensitivity for the model analyte (Fe) in brass samples, and with reasonable precision and accuracy (±15%) for the pulsed laser system used, but with far less sensitivity than might be predicted. Noise studies clearly indicate that laser-induced background fluorescence is the principal limiting noise source. Fundamental material transport studies indicate that diffusional loss of atomic number density to the walls is of much less importance than the background fluorescence in determining the sensitivity of the system. Extrapolations based on cell–experiment design to maximize number density and minimize background emission and fluorescence promise ng g–1 sensitivities for future implementations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.