Abstract

A method for the rapid trace analysis of 17 residual pesticides in water by narrow-bore capillary (I.D. 100 μm) gas chromatography–mass spectrometry (GC–MS) using a programmable temperature vaporizer (PTV) was discussed. The method consisted of a large-volume injection (40 μl) by a PTV, high-speed analysis using a narrow-bore capillary column and MS detection. The PTV with solvent vent mode was very useful for large-volume injection into a narrow-bore capillary column because the injected solvent volume could be reduced to less than 2 μl. The analysis time was 8.5 min [less than 50% of the analysis time using conventional columns (I.D. 250 μm)]. A 10-ml volume of river water was extracted by dichloromethane (4 ml), and then the extract was condensed to 1 ml. This extract was analyzed. Mean recoveries for river water spiked at 100 pg/ml ranged from 83.4 to 96.7%. The limit of detections of the 17 pesticides ranged from 1 to 100 pg/ml.