Abstract

A method for the analysis of trace-level benzalkonium chloride has been established using flow injection ionspray mass spectrometry-mass spectrometry with multiple reaction monitoring. Quantification was carried out using an external standard based on peak area summation of each benzalkonium ion (C8, C10, C12, C14, C16 and C18) in the mass spectra. The multiple reaction monitoring technique provides additional specificity for identification and quantification. The quantification linear dynamic range was found to be 5.0-100.0 ng ml-1, the correlation coefficient > 0.999, and the detection limit 1.2 ng ml-1. The method was applied to quantify benzalkonium chloride on skin, which was sampled with a D-SQUAME tape from skin surface and extracted from the tape with methanol.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.