Abstract
Ultrahigh-performance liquid chromatography coupled with high-field quadrupole Orbitrap high resolution mass spectrometry was used for the separation and determination of 20 antihistamines, and a dispersive micro solid-phase extraction procedure using high-performance absorbing material was developed as a sample preparation strategy for extracting 20 antihistamines from milk. Instrument conditions and key parameters influencing extraction efficiency were investigated to obtain an optimized method. The limit of detection for 20 antihistamines in milk using this method is 0.05 µg/L to 1.0 µg/L. Recoveries are between 80.7 % and 108.3 %, and the relative standard deviation is less than 15 %. It is suitable for confirmatory monitoring and quantitative analysis of 20 antihistamines in milk. The results show that antihistamines in milk may be noteworthy issues for human health and environmental pollution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Journal of chromatography. A
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.