Abstract

Trans-oral robotic surgery (TORS) using the da Vinci surgical robot is a new minimally-invasive surgery method to treat oropharyngeal tumors, but it is a challenging operation. Augmented reality (AR) based on intra-operative ultrasound (US) has the potential to enhance the visualization of the anatomy and cancerous tumors to provide additional tools for decision-making in surgery. We propose a US-guided AR system for TORS, with the transducer placed on the neck for a transcervical view. Firstly, we perform a novel MRI-to-transcervical 3D US registration study, comprising (i) preoperative MRI to preoperative US registration, and (ii) preoperative to intraoperative US registration to account for tissue deformation due to retraction. Secondly, we develop a US-robot calibration method with an optical tracker and demonstrate its use in an AR system that displays anatomy models in the surgeon's console in real-time. Our AR system achieves a projection error from the US to the stereo cameras of 27.14 and 26.03 pixels (image is 540[Formula: see text]960) in a water bath experiment. The average target registration error (TRE) for MRI to 3D US is 8.90mm for the 3D US transducer and 5.85mm for freehand 3D US, and the TRE for pre-intra operative US registration is 7.90mm. We demonstrate the feasibility of each component of the first complete pipeline for MRI-US-robot-patient registration for a proof-of-concept transcervical US-guided AR system for TORS. Our results show that trans-cervical 3D US is a promising technique for TORS image guidance.