Abstract

PTB and DLR join their expertise and experience in optical radiometry and in THz techniques to perform what is to our knowledge the first traceable measurement of radiant power of a THz quantum cascade laser and the first absolute calibration of a THz radiation detector against a cryogenic radiometer (CR). A total standard uncertainty of 7.3% was achieved at a frequency of 2.5 THz corresponding to a wavelength of 120 µm. This uncertainty is dominated by the limited knowledge of the absorptance of the CR cavity. All other uncertainty contributions including those arising from diffraction are only 2%.