Abstract

This preliminary research is focused on the task of defining an equivalent standard column set between cryogenic and flow-modulation comprehensive two-dimensional gas chromatography (GC × GC) combined with mass spectrometry (MS). Cryogenic modulation (CM) was carried out by using a loop-type device, while the flow modulator used was a seven-port wafer chip, equipped with an external accumulation loop.Initially, a common low-polarity + mid-polarity CM GC × GC column set was selected (30 m × 0.25 mm ID × 0.25 μm df + 1.5 m × 0.25 mm ID × 0.25 μm df), a method was developed, and a GC × GC-MS fingerprint was attained (on a sample of bio-oil derived from coconut fibers). After, a column set with the same stationary phases was selected for the flow modulation GC × GC-MS method (20 m × 0.18 mm ID × 0.18 μm df + 5 m × 0.32 mm ID × 0.25 μm df), with the capability to provide a-similar-as-possible separation. A side-by-side measurement of several chromatography parameters (efficiency, peak capacity, resolution, peak widths, retention factors, elution temperatures) was made.