Abstract

Objective. A novel polyacrylamide/polyvinyl alcohol superporous hydrogel (PAAm/PVA SPH)-based semi-dry electrode was constructed for capturing electroencephalogram (EEG) signals at the hairy scalp, showing automatically ‘charge–discharge’ electrolyte concept in EEG electrode development. Approach. In this regard, PAAm/PVA SPH was polymerized in-situ in the hollow electrode cavity by freezing polymerization, which acted as a dynamic reservoir of electrolyte fluid. The SPH can be completely ‘charged’ with electrolyte fluid, such as saline, in just a few seconds and can be ‘discharged’ through a few tiny pillars into the scalp at a desirable rate. In this way, an ideal local skin hydration effect was achieved at electrode–skin contact sites, facilitating the bioelectrical signal pathway and significantly reducing electrode–skin impedance. Moreover, the electrode interface effectively avoids short circuit and inconvenient issues. Main results. The results show that the semi-dry electrode displayed low and stable contact impedance, showing non-polarization properties with low off-set potential and negligible potential drift. The average temporal cross-correlation coefficient between the semi-dry and conventional wet electrodes was 0.941. Frequency spectra also showed almost identical responses with anticipated neural electrophysiology responses. Significance. Considering prominent advantages such as a rapid setup, robust signal, and user-friendliness, the new concept of semi-dry electrodes shows excellent potential in emerging real-life EEG applications.