Abstract

Biopharmaceutical product characterization benefits from the quantification and differentiation of unwanted protein aggregates and silicone oil droplets to support risk assessment and control strategies as part of the development. Flow imaging microscopy is successfully applied to differentiate the two impurities in the size range larger than about 5 µm based on their morphological appearance. In our study we applied the combination of oil-immersion flow imaging microscopy and convolutional neural networks to extend the size range below 5 µm. It allowed to differentiate and quantify heat stressed therapeutic monoclonal antibody aggregates from artificially generated silicone oil droplets with misclassification rates of about 10% in the size range between 0.3 and 5 µm. By comparing the misclassifications across the tested size range, particles in the low submicron size range were particularly difficult to differentiate as their morphological appearance becomes very similar.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.