Abstract

Optimal control of minimally invasive interventions by hyperthermia requires dynamic temperature mapping at high temporal resolution. Based on the temperature-dependent shift of the proton resonance frequency (PRF), this work developed a method for real-time MRI thermometry which relies on highly undersampled radial FLASH MRI sequences with iterative image reconstruction by regularized nonlinear inversion (NLINV). As a first step, the method was validated with use of a temperature phantom and ex vivo organs (swine kidney) subjected to heating by warm water or a pulsed laser source. The temperature maps obtained by real-time PRF MRI demonstrate good accuracy as independently controlled by fiber-optic temperature sensors. Moreover, the dynamic results demonstrate both excellent sensitivity to single laser pulses (20 ms duration, 6 J energy output) and high temporal resolution, i.e., 200 ms acquisition times per temperature map corresponding to a rate of 5 frames per second. In addition, future extensions to in vivo applications were prepared by addressing the breathing-related motion problem by a pre-recorded library of reference images representative of all respiratory states. The proposed method for real-time MRI thermometry now warrants further developments towards in vivo MRI monitoring of thermal interventions in animals.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call