Abstract
The electrohysterogram is a potential new tool for diagnosing preterm labor. Parameters from the electrohysterogram may be influenced by uterine movement. An observational study was performed quantifying uterine movement during labor as a step towards improving electrohysterogram analysis for predicting preterm labor. The uterine wall was continuously tracked by ultrasound imaging during first stage of labor while an accelerometer recorded external abdominal accelerations in six women. A cyclic cranial-caudal movement of the uterine wall, caused by maternal respiration, was observed. This is reported and quantified for the first time. Average frequency, amplitude, and peak speed were 0.27±0.07Hz, 0.68±0.84cm, and 1.04±1.20cm/s, respectively. The accelerometer signal correlated with uterine movement and therefore can possibly provide a reference for removing movement-induced artifacts. There is a need to model and measure the effect of uterine movement on the electrohysterogram parameters and make measurements more robust to movement artifacts.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.