Abstract
We demonstrate ElectroWetting-On-Dielectric (EWOD) transport and SQUID gradiometer detection of magnetic nanoparticles (MNPs) suspended in a 2 microl de-ionized water droplet. This proof-of-concept methodology constitutes the first development step towards a highly sensitive magnetic immunoassay platform with SQUID readout and droplet-based sample handling. Magnetic AC-susceptibility measurements were performed on MNPs with a hydrodynamic diameter of 100 nm using a high-Tc dc Superconducting Quantum Interference Device (SQUID) gradiometer as detector. We observed that the signal amplitude per unit volume is 2.5 times higher for a 2 microl sample droplet compared to a 30 microl sample volume.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.