Abstract

We describe a mini-endoscope design that uses a new type of electrically tunable liquid crystal lens array enabling the dynamic increase of spatial resolution by adjusting the working distance in various zones of interest over a relatively large field of view (FoV) without mechanical movement. The characterization of the system is performed by using uniform fluorescent films, fluorescent micro spheres and a tissue sample expressing the fluorescent calcium indicator GCaMP6s. Lateral resolution of up to 2 µm over the FoV between 300 µm - 400 µm is experimentally demonstrated.