Abstract

Capillary optics are used for X‐ray fluorescence micro‐analysis using the Cu Kα line provided by a rotating anode. The excitation beam is focused using a polycapillary lens on a Co–Ti sample. Cylindrical glass capillaries of various diameters are fitted to the X‐ray detector (Energy Dispersive X‐Ray (EDX) analyzer) and displaced along the irradiated zone of the sample. The fluorescence is studied as a function of capillary position. Good agreement is found between experimental and calculated lateral widths of the fluorescence collection, taken into account the cylindrical capillary critical angles relevant in the experiment. The influence of the cylindrical capillary diameter on the signal level detected is studied to estimate the possibility of lateral resolution increase of X‐ray fluorescence technique both in‐lab and in synchrotron environment. Copyright © 2013 John Wiley & Sons, Ltd.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.