Abstract
Continuous electronic fetal heart rate (FHR) monitoring during labor is motivated by the clinical experience that fetal distress causes loss of FHR variation and the occurrence of decelerations late during uterine contraction. This practice is of uncertain clinical benefit, perhaps because the interpretation is qualitative. We have developed new quantitative measures and analyzed cardiotocograph records from 148 consecutive patients, 44 of whom had at least one "nonreassuring" epoch. In multivariate regression models, measures of deceleration and variability were significantly associated with the obstetrician's diagnosis (receiver operating characteristic area 0.84, p < 0.05). This approach may be useful clinically.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.