Abstract
We demonstrate a portable, compact system to perform absorption-based enzymatic assays at a visible wavelength of 639 nm on a photonic waveguide-based sensor chip, suitable for lab-on-a-chip applications. The photonic design and fabrication of the sensor are described, and a detailed overview of the portable measurement system is presented. In this publication, we use an integrated photonic waveguide-based absorbance sensor to run a full enzymatic assay. An assay to detect creatinine in plasma is simultaneously performed on both the photonic sensor on the portable setup and on a commercial microplate reader for a clinically relevant creatinine concentration range. We observed a high correlation between the measured waveguide propagation loss and the optical density measurement from the plate reader and measured a limit-of-detection of 4.5 μM creatinine in the sensor well, covering the relevant clinical range for creatinine detection.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.