Abstract

The French National Metrology Laboratory LNE-LCM has developed a high temperature reference facility for accurate measurements of the specific heat capacity and of the enthalpy of fusion of materials over the temperature range [23 °C, 1000 °C]. The metrological approach was to modify a commercial Calvet calorimeter in order to lower the uncertainty of measurement and to insure the metrological traceability of the measurements to the SI units, in particular by designing a new calibration system. The enthalpies of fusion of pure metals (indium, tin and silver) and of a binary alloy Ag-28Cu have been measured. The results obtained on the three pure metallic materials are in very good agreement with data obtained by other National Metrology Institutes (NMIs) using adiabatic calorimetry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.