Abstract

AbstractSome lab‐on‐a‐chip applications require to establish a controlled spatial concentration gradient of (chemical) species, for example for iso‐electrical focusing or to study chemotactic properties of cells. We show that covering a microchannel floor with special grooves or ridges, well‐controlled concentration gradients can be created, depending on the geometrical design of the grooves or ridges. In our case, the pattern consists of ridges that are slanted with respect to the main channel direction. Similar patterns have been applied in the past to achieve mixing by introducing chaotic advection. We present experimental and numerical results that prove the mixing effectiveness of the ridges. In addition, making use of the local mixing capabilities of the ridge patterns, we show, using numerical simulations, how to achieve a concentration gradient across a microfluidic channel. magnified image

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.