Abstract
Simultaneous T1 and T2 images have been obtained using multiple-echo self-normalizing sequences. Differential T1 and T2 discrimination for two of these sequences, double saturation recovery and inversion-saturation recovery, have been explored in detail. The inversion-saturation recovery sequence exhibits T1 discrimination comparable to, and in many cases better than, that of optimal T1 weighted conventional pulse sequences, but is poor for T2 imaging. The double saturation recovery sequence yields comparable tissue discrimination for T1 and T2 imaging, but is poor compared to optimal T1 weighted and T2 weighted conventional pulse sequences. Simultaneous T1 and T2 images represent quantitative maps of intrinsic tissue properties, and hence form a natural basis for an automated image analysis and tissue classification system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.