Abstract

This study aimed to investigate improvements in 68Ga-FAPI PET/CT image quality due to using total variation regularized expectation maximization (TVR-EM) and ordered subset expectation maximization (OS-EM) reconstruction. Data from a total of 24 patients were retrospectively analyzed in this study. Positron emission tomography (PET) images were reconstructed using OS-EM and TVR-EM for 2 and 3 minutes-per-bed (min/bed) acquisition. The SUV<inf>mean</inf> of a region-of-interest on the liver, image noise, signal-to-noise ratios (SNR), the SUV<inf>max</inf> of the lesions and the tumour-to-background ratios (TBR) were measured and compared between the 2 methods. Subjective image qualities were evaluated by two experienced radiologists using a 5-point score scale (5-excellent, 1-poor). In total, 132 lesions were analyzed. The image noise in TVR-EM reconstruction groups was lower than in the OS-EM groups (all P<0.05). The SNR, SUV<inf>max</inf> of lesions and the TBR were higher for the TVR-EM reconstruction groups compared to OS-EM groups (all P<0.05). Also, the SUV<inf>max</inf> of the lesions in the TVR-EM groups increased by at least 12% compared to OS-EM 3 min/bed group. The SUV<inf>max</inf> for small (<10 mm) and large lesions (>10 mm) in the TVR-EM 2 min/bed group were significantly larger compared to the OS-EM 3 min/bed groups (all P<0.05). The highest image quality score resulted from the TVR-EM 3 min/bed group with a penalization factor of 0.25 (3.92±0.19). TVR-EM reduces image noise and improved the SNR, SUV<inf>max</inf> and TBR of the lesions. It also enables fast acquisition without compromising image quality compared to standard OS-EM.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call