Abstract

Total sulfur is an analyte for which there are few determinations published, despite the fact that it is a very important element (e.g., a major element in most ores, an important gas constituent in global warming, an active participant in acid drainage). Most geological reference materials have very poor quality sulfur results, that is with relative standard deviations (RSD) in the range of 30–50%, even for concentrations over 100 μg g−1 S, which compromises their use as calibrators. In order to provide modern results with low RSD, sulfur was determined in twenty‐nine geological reference materials with a state‐of‐the‐art elemental S/C analyser using metal chips (certified reference materials with a traceability link) and analytical grade sulfur for high concentration samples. Analytical parameters (sample mass, crucible degassing, calibration strategy, etc.) were optimised by testing. Our results agreed with reference material values provided by issuing bodies. Results for CCRMP SY‐2 (129 ± 13 μg g−1 S), which has been proposed as a sulfur reference material, were in agreement with the proposed modern value of 122 ± 3.7 μg g−1 S.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.