Abstract

To investigate bolus design and VMAT optimization settings for total scalp irradiation. Three silicone bolus designs (flat, hat, and custom) from .decimal were evaluated for adherence to five anthropomorphic head phantoms. Flat bolus was cut from a silicone sheet. Generic hat bolus resembles an elongated swim cap while custom bolus is manufactured by injecting silicone into a 3D printed mold. Bolus placement time was recorded. Air gaps between bolus and scalp were quantified on CT images. The dosimetric effect of air gaps on target coverage was evaluated in a treatment planning study where the scalp was planned to 60Gy in 30 fractions. A noncoplanar VMAT technique based on gEUD penalties was investigated that explored the full range of gEUD alpha values to determine which settings achieve sufficient target coverage while minimizing brain dose. ANOVA and the t-test were used to evaluate statistically significant differences (threshold=0.05). The flat bolus took 32±5.9min to construct and place, which was significantly longer (p<0.001) compared with 0.67±0.2min for the generic hat bolus or 0.53±0.10min for the custom bolus. The air gap volumes were 38±9.3cc, 32±14cc, and 17±7.0cc for the flat, hat, and custom boluses, respectively. While the air gap differences between the flat and custom boluses were significant (p=0.011), there were no significant dosimetric differences in PTV coverage at V57Gy or V60Gy. In the VMAT optimization study, a gEUD alpha of 2 was found to minimize the mean brain dose. Two challenging aspects of total scalp irradiation were investigated: bolus design and plan optimization. Results from this study show opportunities to shorten bolus fabrication time during simulation and create high quality treatment plans using a straightforward VMAT template with simple optimization settings.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call