Abstract

A newly constructed measuring chamber for total reflection X-ray fluorescence analysis (TXRF) has been tested for several applications. Full spectrum excitation and monoenergetic excitation of the sample are compared using rotating and standing anode tubes as primary radiation source. Advantages and disadvantages of using a W-C multilayer structure as monochromator in TXRF are discussed. Extrapolated detection limits of 170 fg (1 fg = 10 −15 g) for Mn were obtained using a Cu-anode and full spectrum excitation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.