Abstract

The parameter variations of exciting radiation, like spectral distribution, intensity, brilliance, polarization and the phenomenon of X-ray total reflection, leads to improved lower limits of detection (LLD) in XRF. Observations and results from experiments performed with different X-ray tubes such as fine focus Cu and Mo anodes, a specially designed Au anode operated with 100 kV and high power rotating anodes are reported. Results from measurements with monochromatic X-rays tuned with a multilayer structure as well as the use of polarized X-rays from the synchrotron will be shown. All developed measuring devices will be described in terms of their recent design features showing the possible geometric arrangements denned by the beam-reflector-detector position. The extrapolated detection limits for the K-shell excitation of rare earth elements are in the region of 0.3 ng, for medium Z elements in the pg range and for optimized conditions, with a rotating Cu anode, 170 fg for Mn are achieved corresponding to the pg g −1 (ppt) concentration level.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.