Abstract

The purpose of this study was to assess a new metal component finishing designed to improve total knee prosthesis durability. Wear of ultrahigh molecular-weight polyethylene (UHMWPE), with generation of submicrometer- and micrometer-sized particles, has been associated with osteolysis and artificial joint failure. Wear extent is influenced by several factors, some of which are related to manufacturing. UHMWPE wear was assessed in metal prosthesis components finished with the Microloy® technology and in traditionally finished components by wear simulation experiments (pin on disk and knee simulator tests) and analysis of wear debris. Microloy®-finished prosthesis showed a 48.5% reduction in UHMWPE total weight loss compared with traditional components (P=0.002). A statistically significant (P<0.05) reduction of UHMWPE debris were detected from the Microloy®-finished compared with the traditionally finished components. These findings suggest the Microloy® metal finishing may enhance the long-term performance of knee prostheses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.