Abstract
We report on the concept, design, and tests of a New Picosecond Event Timer (NPET) device and its response to the total ionizing dose. These radiation tests aimed to verify that the existing version of the NPET device may be used in space missions. The device was subjected to radiation of a total dose of 74 Gy provided by a 60Co source. The epoch timing part of the device was not detectably affected by the radiation dose. After the radiation dose, the epoch timing of single measurement resolution is better than 0.9ps with the overall timing stability characterized by Time Deviation being better than 5fs for averaging times of 100s. The completed device tests and results indicate the possibility of using the existing version of the New Pico Event Timing device electronics for a space mission in which the total radiation dose will not exceed 74 Gy value.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.